Cryoalfa Super Contact, 3, 5 or 7 mm contact cryosurgery
Contact Freezing · 3, 5 or 7 mm gold-plated surface · Class IIa medical device
Diameter
Sale restricted to healthcare and aesthetics professionals qualified to use it, subject to a declaration when ordering.
Description
In the Super Contact the nitrous oxide does not touch the skin. When the lever is pressed, the gas cools from the inside a closed metal head with a gold-plated surface; after about 15 seconds the head reaches about −50 °C and is applied to the lesion with light pressure. The manufacturer calls this method Contact Freezing.
The diameter of the surface is fixed and is chosen at purchase: 3 mm, pointed, for small lesions; 5 mm; 7 mm, flared, for larger surfaces. If you want all three you can take the set. Valve body, lever and cartridges are those of the Cryoalfa Super.
Indications, contraindications and application times are those in the instructions shared by Super and Lux. As for the whole line, the manufacturer restricts it to staff trained in cryosurgery.
Features
- Contact Freezing technique: closed head cooled by N2O
- 3, 5 or 7 mm gold-plated contact surface, not interchangeable
- Ready in about 15 seconds, at about −50 °C
- Lever release, with one finger
- 16 g or 25 g cartridges with valve, removable without leaks
Technical data
- Gas: nitrous oxide, in 16 g (CA-K-V16) or 25 g (CA-K-V25) cartridges with valve
- Head temperature: about −50 °C after about 15 seconds
- Class IIa medical device, notified body 0494; manufacturer codes CA-S-C3, CA-S-C5, CA-S-C7
Included
- Cryoalfa Super Contact in the chosen diameter
- 16 g N2O cartridge
- Rigid carry case with moulded interior
- Instructions for use
Warnings
- For doctors and healthcare staff trained in cryosurgery only
- Clean and disinfect the head no sooner than 10 minutes after use, as stated in the instructions
- Cartridges are under high pressure: never above 50 °C, keep away from sunlight
This product is intended exclusively for healthcare and aesthetics professionals qualified to use it. By purchasing it you declare, under your own responsibility, that you are qualified to use it in your work, and you undertake to follow the manufacturer's instructions and warnings.