Kronos Class B autoclave, 18 litres
Fractionated vacuum · 18 L stainless steel chamber · built-in printer and water check
Internal data logger
OSMO3
Description
A Class B autoclave sterilises everything that passes through a practice: solid and hollow instruments, wrapped or loose, and porous loads such as gauze and fabrics. The fractionated vacuum is what makes this possible: before the steam, air is removed in several stages, so the steam reaches even inside cannulas and into the folds of pouches.
The 18-litre chamber is seamless AISI 304 steel and takes four trays: enough for a day's work in a beauty salon, a podiatry practice or a tattoo studio. The door closes with a motor and stays locked under vacuum until the cycle ends. Every cycle comes out on paper from the built-in printer, ready to file.
Water matters as much as steam: a conductivity meter checks its quality at every load and shows it on the display. With the OSMO3 version the water arrives already purified from a reverse osmosis unit connected to the mains supply, with no canisters to buy and carry; with the internal data logger every cycle is also recorded on a USB stick and archived on the computer.
Features
- Class B with fractionated vacuum
- 10 preset, validated programmes, including the prion cycle at 134 °C
- Helix, Bowie & Dick and vacuum leak test programmes
- Built-in steam generator
- Motorised closing with the door locked under vacuum
- Conductivity meter for water quality
- Built-in thermal printer and multilingual four-button display
- Test result memory (Memory Test System)
- Front pump to fill the tank from a canister on the floor; manual or automatic draining
- Extra drying of 5 or 15 minutes, standby after 60 minutes and delayed start
- Outer panels with antimicrobial treatment
- Connection for an external data logger
Programmes
- Wrapped hollow instruments: 27 min at 134 °C, 40 min at 121 °C
- Wrapped solid instruments: 22 min at 134 °C, 35 min at 121 °C
- Porous loads: 32 min at 134 °C, 45 min at 121 °C
- Prions: 39 min at 134 °C
- Rapid: 19 min at 134 °C, 29 min at 121 °C
- Unwrapped hollow instruments: 20 min at 134 °C
- Helix and Bowie & Dick test: 25 min
- Vacuum leak test: 15 min
Times include preheating and drying and vary with the load and the power supply.
Technical data
- 18-litre chamber, ø 236 × 381.5 mm, AISI 304 steel
- Maximum load: 2.5 kg wrapped, 4 kg unwrapped, 1 kg porous
- 4 aluminium trays, 185 × 285 × 18 mm
- External dimensions 505 × 615 × 400 mm (width × depth × height), weight 52 kg
- Clearance to open the door: 350 mm
- 4-litre tanks for clean and used water
- Power supply 230 V, 50–60 Hz, power 2,400 W
Included
- 4 aluminium trays with tray holder and handle to remove them
- Water fill and drain hoses, sponge
Versions
- Internal data logger: records every cycle on a USB stick, to download to the computer
- OSMO3: separate reverse osmosis unit with carbon pre-filter, membrane and resin filter; connects to the mains water supply, works without electricity and can be placed up to 5 metres from the autoclave
Documents and media
Datasheets, manuals, certifications and videos for this product.